| Hybrid Polaris Spectra® |
Passive Polaris Spectra® |
Polaris Vicra® | |
|---|---|---|---|
| Performance | |||
| Accuracy: | |||
| Volumetric1,2 | Pyramid: 0.25 mm RMS Extended Pyramid: 0.30 mm RMS2 |
0.25 mm RMS | |
| AAK | 0.35 mm RMS | 0.35 mm RMS | |
| 95% Confidence Interval | Pyramid: 0.5 mm Extended Pyramid: 0.6 mm |
0.5 mm | |
| Maximum Update Rate3 | 60 Hz | 20 Hz | |
| Operating Temperature4 | 10°C to 40°C | 10°C to 30°C | |
| Measurement Volume | Pyramid and Extended Pyramid | Vicra Volume | |
| Data Communication Interface | |||
| Interface | USB 1.1 and 2.0 compatible, RS-232 or RS-422 with SCU | USB 1.1 and 2.0 compatible | USB 1.1 and 2.0 compatible |
| Maximum data rate | 1.2 Mb per second | 1.2 Mb per second | 1.2 Mb per second |
| Tools | |||
| Tool Types | Passive, active wireless, active | Passive, active wireless | Passive, active wireless |
| Maximum number of tools | Up to 15 wireless tools (maximum 6 active wireless) with a maximum of 32 passive and 32 active markers in view and up to 9 additional active tools | Up to 15 wireless tools (maximum 6 active wireless) with a maximum of 32 passive and 32 active markers in view | Up to 6 tools (maximum 1 active wireless) with a maximum of 32 passive markers in view |
| Maximum number of markers per tool | 6 single-face/20 multi-face for passive or active wireless tools; 20 for active tools | 6 single-face/20 multi-face for passive or active wireless tools | 6 single-face/20 multi-face for passive or active wireless tools |
| Tool change | Automatic for active; software controlled for
passive and active wireless |
Software controlled | Software controlled |
| Power | |||
| Power Requirements | 100/120/220/240 V AC, ~50/60 Hz, 0.5 A | 100-250 V AC ~50/60 Hz, I A | 100-250 V AC ~50/60 Hz, I A |
| Laser | |||
| Positioning Laser | Class 2 laser Wavelength - 635 nm Maximum output - 1 mW |
Not available | |
| Approvals | |||
| The Polaris System is designed and tested to meet the following electrical safety and electromagnetic compatability standards: | |||
| Electrical Safety | IEC60601-1(1988) Second Edition plus A1:91, A2, A2:95 CAN/CSA-C22.2 No. 601.1-M90, 2005 UL 60601-1, First Edition, 2006-04-26 IEC60825-1:1993 plus A1:1997 and A2:2001 |
||
| Electromagnetic Compatability |
CISPR 11:2004/EN55011:1998 plus A1:1999, A2:2002 FCC Part 15, Class B EN 60601-1-2:2001 JIS T 0601-1-2:2002 |
||
1Based on a single marker stepped through more than 900 positions throughout the measurement volume using the mean of 30 samples at each position at 20°C.
2 Accuracy stated based on overall volume.
3 Varies according to tool combinations with a maximum rate of 60 Hz.
4 If the Host USB Convertor (HUC) is located within the patient vicinity where it may be touched by the patient, the operating ambient temperature range is +10°C to +30°C. If the HUC is located outside this patient vicinity, the operating ambient temperature range permitted for the system is +10°C to +40°C. The power supply unit must be located outside the patient vicinity under all operating conditions.
The measurement volume is the region in which tools can be tracked.
| Polaris Spectra | Polaris Vicra |
|---|---|
![]() (Click for larger image) |
![]() (Click for larger image) |
>> Polaris Spectra and Polaris Vicra Accessories